Contact an Account ManagerSeparation of DNA fragments
Summary
Type: HPLC
Application #: 111460
Catalog#: 736601 - MN HPLC Column (Valco type), Nucleogen DEAE, 4000A, 7um, 6x125mm
Detection: UV, 260 nm, Sensitivity 0.02 AUFS
Conditions: Eluents: A: 50 mM NaH2PO4, 4 M urea, 0.4 M NaCl;
B: A but 1 M NaCl
Gradient:
0 min 55% B
1 min 65% B
1.5 min 65% B
2.5 min 70% B
3 min 70% B
4 min 75% B
4.5 min 75% B
5.5 min 80% B
6 min 80% B
7 min 85% B
7.5 min 85% B
Flow rate: 1.5 ml/min
Authors: LC Department
Source: Macherey-Nagel